The temperature of the Upper Heated Plate is controlled through the instrument firmware. For the best performance the control algorithm requires accurate information concerning the thermal properties of the Upper Heated Plate and the cooling fluid.
In TRIOS software, click the TRIOS button > Options > Discovery HR or AR > Temperature. The window shown below displays.
A list of the features are described below:
Feed | Temperature | Flow Rate | Range |
Mains Tap Water | 15°C | 0.75 Liter min-1 | 5°C |
Fluid Circulator | 5°C | 0.25 Liter min-1 | 15°C |
Vortex Air Cooler | 15°C | ---- | 100°C |
For configurations without an Upper Temperature Sensor (UTS):
For configurations with UTS:
Every UTS is uniquely calibrated with a traceable dry block temperature system during the manufacturing process. These calibration values are stored within the component, and used automatically by the system without operator involvement. The calibration described in Upper Heated Plate (UHP) Zero Heat Flow is not required when using a UTS.